Client Challenge
A biopharmaceutical company developing a lentiviral gene therapy needed a potency assay for its lentiviral vector. Potency was assessed by the reduction of a disease-associated intracellular metabolite in transduced patient-derived cells, relative to untransduced controls. The company needed an LC-MS/MS method that could quantify this metabolite in cell lysates and demonstrate that the percent-reduction readout was repeatable across replicates, analysts, and days.
Avance Biosciences’ Solution
Avance Biosciences qualified a UHPLC-MS/MS method for quantifying the metabolite in cell lysates from transduced and untransduced patient-derived cells. The method included:
- UHPLC separation coupled to a triple quadrupole mass spectrometer with QTRAP for selective, sensitive detection
- Alkylation and derivatization steps in the extraction procedure to prepare the analyte for LC-MS/MS analysis
- Calibration standards, check standards, and blanks bracketing each run to monitor linearity, recovery, and interference
- Normalization of metabolite levels to total protein in each lysate
- Calculation of percent reduction versus untransduced controls as the potency assessment
Qualification Protocol
The study followed a pre-approved protocol with predefined acceptance criteria. Patient-derived cells were transduced with a positive control vector lot and with an independent mock test sample designed to simulate future sample testing, alongside untransduced controls. Nine independent sets, prepared in duplicate by two analysts on two different days, were analyzed across three LC-MS/MS runs.
Assay Qualification
Each parameter addressed a question a quality unit will ask of a potency method:
- Precision: Does the potency readout repeat across replicates, analysts, and days? Repeatability (one analyst) and intermediate precision (two analysts, two days) were both below 9% CV for the positive control and the mock test sample, against a criterion of ≤30%.
- Assay performance: Does the biology behave as expected? In every set, positive control lysates showed lower metabolite levels than untransduced controls, and untransduced replicates met their precision criterion.
- System suitability: Is the instrument fit for use before any result is interpreted? Every run met predefined criteria for chromatographic non-interference, sensitivity, system precision, linearity, and check standard recovery.
Customer Outcome
- A qualified LC-MS/MS method for quantifying the potency of a lentiviral gene therapy vector as percent metabolite reduction
- Documented repeatability and intermediate precision, well within acceptance criteria
- Consistent results across two analysts and two days, with no incidents or deviations
- A signed final report prepared under cGMP requirements
Client Impact
The client now has a qualified, precise LC-MS/MS potency assay ready for use in future potency testing of its lentiviral gene therapy.